Saratov JOURNAL of Medical and Scientific Research

Thermal imaging characteristics of ultrasonic and pneumatic guillotine 25 gauge vitrectors

Year: 2018, volume 14 Issue: №4 Pages: 916-919
Heading: Ophtalmology Article type: Original article
Authors: Aznabaev В.М., Dibaev T.N., Mukhamadeev T.R., Vafiev A.S., Shavaliev I.Kh.
Organization: Bashkir State Medical University of the Russian Health Care, ZAO Optimedservis
Summary:

Purpose: to analyze thermal characteristics of ultrasonic and pneumatic guillotine 25G vitrectors in the experiment. Materials and Methods. In vitro studies were performed using Testo 882 thermal imager under various conditions for a more complete assessment of thermal characteristics of 25G ultrasonic vitrector and 25G pneumatic guillotine vitrector. First experiment was carried out in normal saline solution of room temperature, second experiment was performed in air medium. Results. It was found that during operation of ultrasonic and pneumatic guillotine vitrectors, the temperature threshold of 45°C, which is critical for thermal damage to retinal tissue, was not reached in all experiments. Under severe conditions (working in the air medium without cooling of vitrector tip), different localization of the heating zone in ultrasonic (at proximal part of the needle) and pneumatic (at distal part, near the port) vitrectors was detected. Conclusion. In experimental conditions, the thermal safety of ultrasonic and pneumatic guillotine vitrectors of caliber 25G has been proved by the method of infrared thermal imaging.

Bibliography:
1 Smetankin IG. Temperaturnyye i strukturnyye izmeneniya operatsionnoy rany posle fakoemulsifikatsii. vypolnennoy bimanualnym i koaksialnym metodami. Meditsinskiy Almanakh 2009; 3 (8): 126-9
2 Young КН, Kevin MM. Heat production: Longitudinal versus torsional phacoemulsification. J Cataract Refract Surg 2009; (35): 1799-805
3 Ligget PE, Prince KJ, Astrahan M, et al. Localized current field hyperthermia: effect on normal ocular tissue. International Journal of Hyperthermia 1990; (6): 517-27.4 Yagura K, Shinoda K, Matsumoto S, et al. Intraoperative Electroretinograms before and after core vitrectomy. PLoS One 2016; (3): 1-10
5 Landers MB, Watson JS, Ulrich JN, et al. Determination of retinal and vitreous temperature in vitrectomy. Retina 2012; 32(1): 172-6
6 Romano MR, Vallejo-Garcia JL, Castellani C, et al. The effect of temperature changes in vitreoretinal surgery. Transl Vis SciTechnol 2016; 5(1): 1-10
7 Iguchi Y, Asami T, Ueno S, et al. Changes in vitreous temperature during intravitreal surgery. Investigative Ophthalmology & Visual Science 2014; (55): 2344-9
8 Pastor-ldoate S, Bonshek R, Irion L, et al. Ultrastructural and histopathologic findings after pars plana vitrectomy with a new hypersonic vitrector system. Qualitative preliminary assessment. PLoS One 2017; 12 (4): 1-16
9 Aznabayev BM, Dibayev Tl, Mukhamadeyev TR, et al. Eksperimentalnoye issledovaniye proizvoditelnosti ultrazvukovogo vitreotoma. Kataraktalnaya i refraktsionnaya khirurgiya 2017; 2 (17): 48-51
10 Martinez-Castillo V, Verdugo A, Boixadera A, et al. Management of inferior breaks in pseudophakic rhegmatogenous retinal detachment with pars plana vitrectomy and air. Arch Ophthalmol 2005; 123 (8): 1078-81
11 000 «Testo Rus». Teplovizor Testo 882: Rukovodstvo polzovatelya. URL: https://www.testo.ru/ru- RU/testo-882/p/0560-0882 (24 March 2018).

AttachmentSize
2018_4-2_916-919.pdf716.51 KB

No votes yet